IOLMaster — Data Management.
Transmitting/exporting data (optional feature)
Patient data can be exported to
- connected office administration systems or personal computers
- a USB storage medium or a CD-RW.
The data is then available for further processing.
 |
NOTE: Transmitting/exporting does not work in the Patient Manager,
only in measurement modes! |
Exporting data to another system
- Data can only be exported to office management systems by the
supplier of such systems. Please contact the respective supplier.
- The appropriate accessories are required for exporting to a
connected Windows-based personal computer. These can be
obtained from Carl Zeiss Meditec. They include a serial cable (null
modem, female/female connector) and software (on CD ROM) to be
installed on the PC. Data is imported to a database on the PC. From
there, data can be exported to other file formats. The graphs of axial
length measurements are made available in JPEG format.
 |
- To export data press the <S> key (not in Patient Manager!) or the
SEND button.
The data will be exported.
|
 |
NOTE: The PC must have been switched on and the software for data
receipt started. A progress bar will be visible on the screen of the
IOLMaster. Data can be archived on the PC or processed in the
appropriate form. |
The export of measured values depends on whether the additional
software Option A plus is installed:
- Without Option A plus: only the measured values and the marked
IOL will be exported.
- With Option A plus the measured values and all calculated lenses
will be exported (see User Database), depending on the setting in
Program Settings/Export.
Exporting data to a storage medium
- Select the desired storage medium in the menu Options - Setup -
Program settings/Export (see User Database).
 |
NOTE: If you wish to export to a CD-RW, you must insert a formatted
CD-RW into the drive. The CD-RW must be formatted elsewhere
(e.g. office PC) in UDF format. Alternatively, use one of the
formatted CD-RWs as supplied. For exporting to an USB flash
drive the latter should enable at least a transfer rate to USB-1.1. |
 |
- To export data to a USB storage medium or a CD-RW press the <X>
key or the EXPORT icon.
|
Data will be available in a text file conforming to the export settings (see