IOL Power Calculations

  IOL Power Calculations
Physician Downloads
Cataract Surgery Arizona
 
IOL Power Calculations IOL Power Calculations
Zeiss IOLMaster Zeiss IOLMaster
Contact Us
Disclaimer
Physician Downloads Surgically-induced Astigmatism
Surgically-induced Astigmatism
Haigis Formula Optimization
Haigis Instructions
Refractive Vergence Formula
4-Formula IOL Constants
Vertex Distance
IOLMaster Checklist (pdf)
IOL Master Manual Download IOLMaster Manual V.5 (pdf)
IOL Master Manual Download IOLMaster Manual V.4 (pdf)
IOL Master Manual Download IOLMaster Manual V.3 (pdf)
Links
References
Special Thanks

Physician Downloads.


Clicking on any download link below will cause the item
to automatically download onto your computer.

Surgically Induced Astigmatism (SIA) Calculator.

Surgically Induced Astigmatism
        Calculator SIA Calculator — SIAC_104.xls (Microsoft Excel)

Surgically Induced Astigmatism
        Calculator Instructions SIA Calculator Instructions — SIA_Instructions.doc (Microsoft Word)

This Surgically Induced Astigmatism Calculator (SIA Calculator) has been designed to calculate, by means of vector analysis, the amount of surgically induced astigmatism created during the cataract surgical procedure. Please feel free to download this calculator, place it on your computer and then add cases sequentially.

Windows: Microsoft Office 200, or later is required. In order to run the SIA Calculator application the Macros must be enabled. (Choose Tools> Macro > Security and choose Medium for a signed macro.)

Macintosh: Office 2001 for Mac or later is required and Macros must be enabled. Detailed instructions can be found under the instructions tab of the SIA Calculator download.

When entering patient data into the SIA Calculator, it is always best to use the same instrument for pre and post-op keratometry measurements. Manual keratometry or autokeratometry is preferred over simulated Ks. Also, please take care not to flip the pre and post-op meridians. Pre and post-op data should be entered in the corresponding steep or flat meridian. After cataract surgery, the axis in each of the two principal meridians should only shift between 5 degrees and 20 degrees; and, unless a very large incision is used, the power change should be less than 1.00 D.

Keratometry measurements are typically between 35.00 and 50.00 diopters. Measurements outside this range are unusual and should be re-verified. Axis measurements must be an integer between zero and 180 degrees. The axis of the steepest and flattest meridian should differ by 90 degrees +/- 5 degrees. Values outside this range may indicate irregular astigmatism and should be re-verified. The incision location must be an integer between zero and 360 degrees. Look at the picture under the SIAReport tab for a graphic representation of the convention for incision location.

This individual patient information will be automatically summarized in the SIA Calculator by incision type, size, and location by using the built-in "Create Report" function.

After you have enough cases to generate a report, the magnitude of surgically induced astigmatism will be visible under the SIAReport tab. This information can then be entered into the "Surgically Induced Cylinder (SIC)" field of the Alcon AcrySof Toric IOL Calculator (acrysoftoriccalculator.com/Country.aspx).

Cataract Surgery Arizona
A tutorial for its use can be viewed by clicking on the "Tutorial" menu item at the bottom of the calculator page.
Cataract Surgery Arizona
Alcon AcrySof Toric
    IOL Calculator PLEASE NOTE

Dr. Hill requests your assistance in putting together a database of information collected from surgeons around the world to be included in an analysis of surgically induced astigmatism. These data will be analyzed in total, so that we may further understand surgically induced astigmatism. If you would like your data to be included in this important analysis, please upload your SIA Calculator information to the database. Instructions for sending data can be found under section VII (Send Data). Please note that no patient names will be sent to the database.

Intraocular Lens Power Calculations
TOP

Haigis Formula Optimization.

Haigis Formula Haigis Formula Optimization — Haigis-300.xls (Microsoft Excel)

This Excel spreadsheet can be used to submit data for optimizing the a0, a1 and a2 constants for the Haigis formula. The corneal power should be entered in diopters and not mm.

Haigis Formula Optimization IMPORTANT:
Because this exercise requires Dr. Hill a significant time to complete, requests for this service must come DIRECTLY FROM THE SURGEON whose data this represents and NOT from support staff. Requests from support staff WILL BE RETURNED UNANSWERED.

Address your subject line to read: "Haigis optimization." Along with the completed Haigis-300.xls spreadsheet, the text of the e-mail should include the manufacturer and model of the IOL, such as Alcon SN60WF, the biometry method employed, such as the IOLMaster and the name of the requesting surgeon. If less than 200 valid cases are submitted, only the a0 Haigis constant will be optimized with the Haigis a1 and the a2 constants remaining at default values. It is best to submit approximately 225 cases so that more than 200 cases will remain after the exclusion of any outliers.

 


 

Haigis Formula Optimization ~ Instructions.

iol power calculations Before entering data into your Excel spreadsheets, please download
      these instructions (below) and follow them exactly.

iol power formula calculations Haigis-instructions.doc (Microsoft Word file):

This download is a set of detailed instructions on how to complete and submit the Excel spreadsheet that will be used to derive your set of surgeon and IOL specific a0, a1 and a2 Haigis constants.

 


 

Refractive Vergence Formula.

Refractive Vergence Formula Refractive Vergence Formula — R-verg-Hill.xls (Microsoft Excel)

A Microsoft Excel spreadsheet containing the refractive vergence formula as outlined in:

Holladay JT: Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 1997;23:1356-1370.

The more sophisticated "Holladay R Formula" is recommended and is part of the Holladay IOL Consultant software package.

Intraocular Lens Power Calculations
TOP

Four-Formula IOL Constants Spreadsheet.

IOL Constants Four-Formula IOL Constants — 4-IOL-constants.xls (Microsoft Excel)

This Microsoft Excel spreadsheet can be used to calculate an average IOL constant based on data from 100 patients for four popular formulas (Holladay 1, Hoffer Q, SRK II and SRK/T). Data must be COMPLETE for all 100 patients entered into the spreadsheet for it to work correctly.

This spreadsheet can be used to arrive at an IOL constant starting point prior to formal optimization using the Holladay IOL Consultant, or the internal lens constant optimization software that comes with your biometer (standard A-scan ultrasound, or the IOLMaster).

This spreadsheet should NOT to be used as the sole method of IOL constant optimization. If you have less than 100 cases entered, you may send the spreadsheet to Dr. Hill for analysis at hill@doctor-hill.com.

 


 

Correcting for Vertex Distance.

IOL Constants Correcting for Vertex Distance — Vertex.xls (Microsoft Excel)

Be sure to maintain the correct sign (+ or -) when entering the refractive data, or the results will not be accurate.

 


 

IOLMaster Checklist.

IOL Constants IOLMaster Checklist — IOLMaster_checklist.pdf (Adobe Acrobat PDF)

The above .pdf file contains the checklist that our staff uses when measuring the axial length, central corneal power, ACD, and horizontal corneal diameter using the IOLMaster. The .pdf file contains three identical checklists. We suggest separating each copy and having them laminated individually to allow various staff members easy access to their own copy.

 

IOL constant optimization

Intraocular Lens Power Calculations
TOP

Intraocular Lens Power Calculations

Zeiss IOLMaster User Manual


East Valley Ophthalmology provides this on-line information for educational and communication purposes only and it should not be construed as personal medical advice. Information published on this website is not intended to replace, supplant, or augment a consultation with an eye care professional regarding the viewer/user's own medical care. East Valley Ophthalmology's disclaims any and all liability for injury or other damages that could result from use of the information obtained from this site. Please read our full Disclaimer.