Lenstar Lens Constants

It is important to remember that these numbers simply represent a starting point, with optimization to follow.

These lens constants are not meant to be definitive.

To begin the process of optimization, please visit our Calculators & Downloads.

IOL Model Mfr A-constant Haigis a0, a1 & a2 Holladay-1 Surgeon Factor Hoffer-Q pACD SRK/T A-constant Implants Contributors
Alcon MA60AC 118.4 a0 = 0.585
a1 = 0.221
a2 = 0.151
1.64 5.39 118.71 220 11
Alcon SN60WF 118.7 a0 = -0.576
a1 = 0.311
a2 = 0.197
1.81 5.61 119.02 2111 1, 2, 3, 4, 5, 6, 9, 10, 11, 13
AMO Tecnis
Z9002
118.7 a0 = -1.185
a1 = 0.211
a2 = 0.244
1.97 5.78 119.20 289 8
AMO Tecnis
ZMB00
118.8 a0 = -0.735
a1 = 0.293
a2 = 0.219
2.11 5.90 119.47 231 12
AMO Tecnis 1 ZCB00 118.8 a0 = -0.028
a1 = 0.313
a2 = 0.184
2.07 5.88 119.37 208 7
B&L Akreos
Adapt AO
118.0 a0 = -0.418
a1 = 0.259
a2 = 0.186
1.60 5.39 118.63 222 9

 

Physician Contributors

  1. Douglas Koch, MD, Houston, TX, USA
  2. Li Wang MD PhD, Houston, TX, USA
  3. Johnny Gayton, MD, Robbins-Warner, GA, USA
  4. Sheridan Lam MD, Chicago, IL, USA
  5. Warren Hill MD, Mesa, AZ, USA
  6. Les Grosinger MD, Bloomfield Hills, MI, USA
  7. Joseph Sokol MD, Shelton CT, USA
  8. Aaron Kuzon MD, Mount Vernon, VA, USA
  9. Brad McCuaig MD, Kitchener, ON, CANADA
  10. John Zdral MF, Fullerton, CA, USA
  11. Nitin Patel MD, Navsari, Gujarat, INDIA
  12. Joseph Bishop, Hilton Head Island, SC, USA
  13. Randy Ennen, MD, Ft. Smith, AR, USA

 

It is important to remember that these numbers simply represent a starting point, with optimization to follow.

These lens constants are not meant to be definitive.

To begin the process of optimization, please visit our Calculators & Downloads.